仪器使用视频  
界面化学基础原理  
表面/界面张力测试技术及其应用  
接触角测试技术及其应用  
界面化学参考文献资料  
界面流变及动态表面/界面张力分析  
精密机械原理  
TOF相机  
新品展示
关于我们
新闻中心
  产品展示
 
  联系我们
 
 
  > 首页 > 技术文章
几家不同公司的接触角仪软件采用停滴法并用Young-Laplace方程拟合测试接触角的区别

本视频中演示了采用Young-Laplace方程拟合时各家接触角仪公司软件的区别。从测试结果来看,接触角变化范围从103-105度之间,属于一个合理的变化范围。这种变化由各种因素造成,如水的挥发等。

 

 

另可参考如下资料,我们发现国外的部分生产厂的接触角仪软件使用Young-Laplace拟合时,90-140度时,测值偏大5度,而大于140度时,测值偏大约10度。

We measured contact angle of same sample with 4 different software from Kruss, Dataphysics, Biolin and KINO. And we measured contact angle of sample by Biolin's sofware - OneAttension V2.4 at first because only this software cannot import images into software and calculate contact angle for imported image. Other sofwares such as DSA1.92 and DSA3.0 from Kruss, and SCA202 from Dataphysics and CAST3.0 from KINO can all import image into software and can calculate contact angle using different method such as circle fitting method, ellispe fitting method and Young-Laplace equation fitting method.

At this article, we just show you the calculated result by different company's software. Comparison of measured result maybe has some existed deviation due to our misoperation. You can contact us to obtain the recorded video for measuring process of contact angle.

And from the comparison of measured contact angle from different drop shape software, may safely draw a conclusion that:

1, When contact angle below 3°:  anti-dazzling screen must be used to get sharp drop shape and when using Young-Laplace equation fitting method, only sofware from Biolin (OneAttension V2.4) is failure to fitting the drop shape and get a acceptable contact angle.

2, When there exists some noise points or poor backgroud, tangent method based on parts of drop profile will  failure to fitting the drop shape and get a acceptable contact angle.

3, When contact angle between 90°-140°, calculated contact angle based on Young-Laplace equation fitting method from drop shape analysis software from Kruss, Dataphysics and Biolin will be larger than other mehod such as circle fitting and ellispe fitting about 3-5° when circle fitting or ellispe fitting can fitting the drop shape. And also this value based on Young-Laplace equation fitting method is larger than ADSA-RealDrop based method from KINO about 3-5°.

4, When contact angle larger than 140°,  calculated contact angle based on Young-Laplace equation fitting method from drop shape analysis software from Kruss, Dataphysics and Biolin will be larger than other mehod such as circle fitting and ellispe fitting about 8-10° when circle fitting or ellispe fitting can fitting the drop shape. And also this value based on Young-Laplace equation fitting method is larger than ADSA-RealDrop based method from KINO about 8-10°.

5, Calculated ontact angles by different method from KINO's sofware CAST3.0 are almost kept same when both of them fitting drop shape correctly.

 Please be noted that,

1, A basic principle is contact angle is calculated from tangent at contact point or curve radius at contact point. And when fitting drop profile with some equation of curve, if result of fitting is good, calculated contact angle will be almost same. In other words, Young-Laplace equation can be used to calculate contact angle when drop shape cannot fitting with circle equation due to effect of gravity. But if we used other equation of curve and fitting it correctly, the result will be kept almost same.

2, Red number means it is failure to fitting drop profile with current method. And "—" means current method cannot calculate any result automatically.  

Comparison of calculated contact angle based on different method and different company such as Kruss, Dataphysics, Biolin and KINO.

   
首页 / 新闻中心 / 产品展示 / 技术文章 / 下载中心 / 关于我们 / 客户留言 / 人才招聘 / 联系我们

版权所有 2005-2015 上海梭伦信息科技有限公司 电话:021-51036075,400 0055117,
上海梭伦信息科技有限公司(www.solon17.com)主要提供:接触角测试仪 ,表面张力仪,界面张力仪,接触角测量仪,
水滴角测试仪,水滴角测试仪接触角测定仪,光学接触角测量仪
网站地图 站点地图

相关搜索:接触角测定仪原理水滴角测量仪器操作步骤接触角测试仪使用方法接触角测试仪操作规程水滴角测试仪详细使用方法水滴角测试仪详细使用方法接触角测量仪的结构界面张力仪使用注意事项表面张力仪的清洁与否对所测数据有何影响?表面张力仪接触角仪表面张力测试仪使用方法表面张力测量仪